New Artisse intrasaccular device for intracranial aneurysm treatment: short term clinical and angiographic result from the prospective registry INSPIRE-A.
No Thumbnail Available
All Authors
Hohenstatt, S.
Costalat, V.
Dargazanli, C.
Killer-Oberpfalzer, M.
Schreiber, B.
Rautio, R.
Sinisalo, M.
Lamin, S.
Chew, HS.
Spelle, L.
LTHT Author
Patankar, Tufail
LTHT Department
Radiology
Interventional Neuroradiology
Interventional Neuroradiology
Non Medic
Publication Date
2025
Item Type
Journal Article
Language
Subject
Subject Headings
Abstract
BACKGROUND: Intrasaccular devices have broadened treatment options for wide necked aneurysms. This study presents the preliminary experience with the Artisse 2.0 device.
METHODS: Innovative NeurovaScular Product SurveIllance REgistry (INSPIRE) is a non-randomized, multicenter, real world clinical study with treatment arms for aneurysms (INSPIRE-A) and acute ischemic stroke (INSPIRE-S). This interim analysis included 87 patients enrolled from November 2022 to April 2024 in the INSPIRE-A Artisse cohort across 16 Europoean centers. Procedures followed standard clinical care, with 6 months of follow-up. Safety and efficacy endpoints included major stroke, neurological death, serious adverse events (SAEs), aneurysm occlusion, and retreatment rates. An independent core laboratory assessed imaging, and all SAEs were reviewed by a clinical events committee. The Artisse steering committee provided independent oversight of the data.
RESULTS: The Artisse device achieved an overall successful implantation rate of 96.6% (84/87), with satisfactory placement rates of 98.7% (74/75) for unruptured and 88.9% (8/9) for ruptured aneurysms. Following the procedure, 46.2% of unruptured aneurysm patients were receiving antiplatelet therapy (APT), predominantly aspirin monotherapy, while no ruptured aneurysm patients received APT. Device related SAE rate was 1.3% (1/87), and the overall stroke rate was 2.3% (2/87), including both ruptured and unruptured aneurysms. At 6 months, 80.0% (28/35) of patients with unruptured aneurysms showed complete obliteration, with no recurrences or retreatments.
CONCLUSIONS: Preliminary experience with the Artisse 2.0 device demonstrated high technical success, favorable safety, and efficacy in aneurysm obliteration at 6 months. Larger studies with longer follow-up periods are needed to confirm these findings.